Pixel Count‚ H x V |
4.2 MPixel‚ 2048 x 2048 |
Optical Format |
1″ |
Image Area |
11.3 x 11.3 mm |
Pixel Dimension |
5.5 x 5.5 µm |
Min. Beam (10 pixels) |
~55 µm |
Shutter Type |
Global |
Max Frame Rate* |
60 Hz |
Single Pulse Capture PRR |
USB 2.0: 6.3 kHz
USB 3.0: 12.6 kHz |
Signal to RMS Noise |
2,500:1 (34 dB optical / 68 dB electrical) |
Electronic Shutter Range |
USB 2.0: 12,600:1 (41 dB)
USB 3.0: 25,000:1 (44 dB) |
ADC |
12-bit |
Measurable Sources |
CW beams‚ pulsed sources; CW to 12.6 kHz with single pulse isolation
Software configurable Auto-trigger‚ Synchronous & Variable Delay |
Measured Beam Powers |
See Charts |
Manual Beam Attenuation |
Includes ND-1, ND-2, and ND-4 magnetic ND filters or C-mount ND filters |
Displayed Profiles |
Line‚ 2D & 3D plots. Normalized or un-normalized.
Linear or Logarithmic‚ Zoom x10
2D‚ 3D in 10‚ 32 or max. colors or grayscale
Contoured display at 10 and 32 colors |
Measured & Displayed Profile Parameters |
Raw and smoothed profiles
Triangular running average filter up to 10% FWHM |
Beam Diameter |
Diameter at two user set Clip levels
Gaussian & ISO 11146 Second Moment beam diameters
Equivalent diameter above a user defined Clip level
Equivalent Slit and Knife Edge diameters |
Beam Fits |
Gaussian & Top Hat profile fit & % fit
Equivalent Slit profile |
Beam Ellipticity |
Major‚ Minor & Mean diameters. Auto-orientation of axes. |
Centroid Position |
Relative and absolute
Intensity Weighted Centroid and Geometric Center
Beam Wander Display and Statistics |
Measurement Accuracy (not limited to pixel size) |
0.1 µm processing resolution for interpolated diameters.
Absolute accuracy is beam profile dependent – ~1 µm accuracy is frequently achievable.
Centroid accuracy is also beam dependent (as good as ±1 µ since it is arithmetically determined from all pixels above the centroid clip level). |
Processing Options |
Image & profile Averaging‚ 1‚ 5‚ 10‚ 20‚ Continuous.
Background Capture and Subtraction.
User set rectangular Capture Block for capture
User set or Auto ellipse Inclusion region with beam tracking for processing
*.ojf files save all WinCamD custom settings for particular test configurations |
Pass/Fail display |
On-screen selectable Pass/Fail colors. Ideal for QA & Production. |
Log Data and Statistics |
Min.‚ Max.‚ Mean‚ Standard Deviation‚ to 4096 samples |
Relative Power Measurement |
Rolling histogram based on user’s initial input. Units of mW‚ µJ‚ dBm‚ % or user choice (relative to a reference measurement input) |
Fluence |
Fluence‚ within user defined area |
Certification |
RoHS‚ WEEE‚ CE |
Multiple Cameras |
Up to 4 cameras, parallel capture.
1 to 8 cameras, serial capture. |
Head Dimensions‚ Width x Height x Depth |
1.8 x 1.8 x 0.8” (46 x 46 x 20 mm) |
Optical depth from housing/filter to sensor (no window) ±0.2 mm |
7.5 mm (12.8 mm with MagND filter) |
Mounting |
8-32 thread, 8 mm deep |
Weight‚ Camera w/ MagND and filter cover |
3.75 oz (106 g) |
Minimum PC Requirements |
Windows 7/8/8.1/10 64-bit, 4 GB RAM, USB 2.0/3.0 port |